VERTICAL LAMINAR DOWNFLOW BIOLOGICAL SAFETY CABINETS
Teknopak offers a wide range of Laminar Airflow Biological Safety Cabinets conforming to current ISO/TC, NSF/CDC/NIH standards and guidelines. Employing contemporary technologies in microcontamination control, these systems effectively provide both bioexclusion and biocontainment.
Here we describe the design that is in common use in the country. However, we do undertake manufacture of other categories against specific orders, including special design innovations for safe exhaust and safe removal of contaminated filters.
All Teknopak units bear the hallmark of committed professionalism, and promise you quality, security, protection and full-value-for-money. Backed by rich experience, continuing research and strong infrastructure, Teknopak’s cost-performance-benefit equations remain unequaled.
Choose Teknopak; and you can be sure you are investing in the best in the market.

|
VERTICAL LAMINAR DOWNFLOW BIOLOGICAL SAFETY CABINETS |
||||||
|
Model |
Work space dimensions |
Overall dimensions |
||||
|
w |
d |
h |
W |
D |
H |
|
|
2293-SM |
36" |
24" |
32" |
40" |
32" |
94" |
|
2294-SM |
48" |
24" |
32" |
52" |
32" |
94" |
|
2296-SM |
72" |
24" |
32" |
76" |
32" |
94" |
|
Execution |
||
|
Standard execution |
Optional features |
Typical applications |
|
|
|
|
Special Components |
|
|
Prefilter |
Washable, reusable fire-retardant depth filter, specially treated to inhibit microfloral growth |
|
HEPA filter |
Microfibreglass media. Efficiency > 99.97% down to 0.3 m at both 100% & 20% of rated airflow |
|
Impellers |
Statically and dynamically balanced for high performance, low noise and vibration |
|
Blower-motor |
Direct-driven dual outlet design with special sealed bearings and unique suspension |

CATEGORIES OF VERTICAL LAMINAR DOWNFLOW BIOLOGICAL SAFETY CABINETS
The Class I BSC

A. front opening
B. sash
C. exhaust HEPA
D. exhaust plenum
The Class II, Type A BSC

A. front opening
B. sash
C. exhaust HEPA filter
D. rear plenum
E. supply HEPA filter
F. blower
|
The Class II, Type B1 BSC (classic design). Connection to building exhaust system required.
A. front opening Note: The cabinet exhaust needs to be connected to the building exhaust. |
The Class II, Type B1 BSC (bench top design). Connection to building exhaust system required.
A. front opening Note: The cabinet exhaust needs to be connected to the building exhaust. |
The Class II, Type B2 BSC.

A. front opening
B. sash
C. exhaust HEPA filter
D. supply HEPA filter
E. negative pressure exhaust plenum
F. supply blower G. filter screen
Note:The carbon filter in the building exhaust is not shown. The cabinet exhaust needs to be connected to the building exhaust system.
The tabletop model of a Class II, Type B3 BSC.

A. front opening
B. sash
C. exhaust HEPA filter
D. supply HEPA filter
E. positive pressure plenum
F. negative pressure plenum
Note: The cabinet exhaust needs to be connected to the building exhaust system.
The Class III BSC.

A. glove ports with O-ring for attaching arm-length gloves to cabinet
B. sash
C. exhaust HEPA filter
D. supply HEPA filter
E. double-ended autoclave or pass-through box
Note: A chemical dunk tank may be installed which would be located beneath the work surface of the BSC with access from above. The cabinet exhaust needs to be connected to the building exhaust system.

Thimble unit for ducting a Class II, Type A BSC.

Note: There is a 1" gap between the thimble unit (D) and the exhaust filter housing (C), through which room air is exhausted.
A. balancing damper
B. flexible connect to exhaust system
C. cabinet exhaust HEPA filter housing
D. thimble unit
E. BSC
A bag-in-out filter enclosure allows for the removal of the contaminated filter without worker exposure.

A. filters
B. bags
C. safety straps
D. cinching straps
E. shock cord located in the mouth of the PVC bag restricts the bag around the second rib of the housing lip.

|
Tech-brief |
|
|
Governing standards |
ISO 14644 (1998), ISO 14698, NSF 49, BS 5726 Appendix A |
|
Air velocity |
90 fpm (0.45 mps) 20% |
|
Cleanliness level |
Fewer than 100 particles of size 0.5 m or larger per cu.ft of air sampled ( or <3,520/cuM ) |
|
Noise level |
< 65 dB (A Scale) |
|
Vibration level |
< 2.5 m (0.0001") |
|
Illumination level |
800 lumens/sq m (75 ft-candles) at work surface |
|
Electricals |
230V AC, 1 f, 50 Hz |
![]() |